Send Message
Home > products > POCT Test Kit >
Diagnostic Kit For Serum Ferritin Immunochromatographic Assay By Wwhs

Diagnostic Kit For Serum Ferritin Immunochromatographic Assay By Wwhs

Hematology serum ferritin test kit

Phosphate buffer serum ferritin test kit

NIR-1000 serum ferritin test kit

Place of Origin:

CN

Brand Name:

WWHS

Certification:

ISO13485

Model Number:

Ferr

Contact Us

Request A Quote
Product Details
Sample Buffer:
25 Vials
Other Usage:
Hematology, Oncology
Preponderance:
Convenience,accurate
Manufacturer:
WWHS Biotech INC
Reactivity:
Human
Type:
Nutritional Metabolism
Payment & Shipping Terms
Minimum Order Quantity
Negotiate
Price
Negotiable
Packaging Details
Colorful Paper Box
Delivery Time
Two weeks
Payment Terms
L/C, T/T, Western Union
Supply Ability
50000 Kits per Week
Product Description

Diagnostic Kit for Serum Ferritin(Immunochromatographic assay)
 

Ⅰ.Product Description

Name Quantity Component
Test card 25/40 It is composed of fluorescent pad (coated with fluorescent labeled Ferr monoclonal antibody), nitrocellulose membrane (coated with Ferr monoclonal antibody and Goat anti mouse IgG antibody), absorbent paper and backing
Sample buffer 25/40 Phosphate buffer
ID card 1 Record the standard curve information of this kit
The components in different batches of kits cannot be used interchangeably.

Package specification:25Tests/kit,40 Tests/kit
Intended use:Diagnostic Kit for Serum Ferritin is used to quantitatively detect the content of ferritin in human serum. Clinically, it is mainly used for the auxiliary diagnosis of iron metabolism related diseases, such as hemochromatosis and iron deficiency anemia.
Test principle:The principle of immunofluorescence chromatography was applied to the Diagnostic Kit for Serum Ferritin. The Ferr antigen in the sample was first bound with the conjugated compound of fluorescent labeled Ferr monoclonal antibody, then moved and combined with another Ferr monoclonal antibody fixed on the nitrocellulose membrane, and the double antibody sandwich complex was formed at the detection line of the cellulose nitrate membrane.
 
Ⅱ.Using the step
1. Before the test, please read the instructions completely. If the test card and sample are stored in cold storage, they should be balanced at room temperature (15-30)℃ for not less than 30min before use.
2. Start NIR-1000 dry fluoroimmunoassay analyser and correctly select the corresponding sample type on the instrument.
3. Take out the ID card, make sure that the batch number of the ID card is consistent with that of the test card, and insert the ID card into the ID card port of the instrument.
4. Take out the test card from the aluminum foil bag and use it within 15 minutes.
5. Place the test card on a clean horizontal table and mark it horizontally.
6. Mix 20 µL of sample with 400µL of sample diluent. Apply 100 µL of diluted samples to the well of the test card.
7. At 5 minutes after addition of samples, insert the test card into NIR-1000 dry fluoroimmunoassay analyser and click the “Instant test” button to read the results.
 
 
Ⅲ.Reference interval

Through the determination of serum samples from 301 healthy people aged 10 ~ 87, the results were statistically analyzed.

Gender

2.5th percentile

(ng/mL)

97.5th percentile

(ng/mL)

Male 24 335
Female 11 307

 
Ⅳ.Why to choose us?

1. Analysis sensitivity
No higher than 5ng/mL.
2. Accuracy
The relative deviation from the target value is within ±15%.
3. Precision
The within and between assay coefficient of variations are within 15%.
4. Linear range
Within the linear range (5 ~ 500ng/mL), the linear correlation coefficient R≥0.990

 
Ⅴ.FERR.pdf
Diagnostic Kit For Serum Ferritin Immunochromatographic Assay By Wwhs 0Diagnostic Kit For Serum Ferritin Immunochromatographic Assay By Wwhs 1

Send your inquiry directly to us

Privacy Policy China Good Quality POCT Test Kit Supplier. Copyright © 2021-2024 fiapoct.com . All Rights Reserved.